STI Publications - View Publication Form #14160
Back to Search Results |
Publication Information
Title | A Prototype Gamma Tomosynthesis System for Molecular Breast Imaging | ||||
Abstract | We are developing an improved gamma camera for molecular breast imaging (MBI) that allows tomographic imaging and improved signal-to-noise over conventional planar methods. We have built a novel gamma camera for MBI that utilizes limited angle tomography, similar to x-ray tomosynthesis, to generate images with depth information and improved signal-to-noise. A unique feature of the gamma tomosynthesis system is a variable-angle slant hole (VASH) collimator. This collimator allows the camera to remain flush against the compression paddle during the tomographic acquisition, which achieves high spatial resolution and simplified detector motion. We constructed a VASH collimator from a stack of tungsten sheets, each sheet containing a matrix of holes created by photo-etching. With the holes of the sheets aligned, a parallel-hole collimator is created; shearing the sheets creates a variable angle slant hole collimator. The shearing is controlled by a motorized mechanism. The collimator was | ||||
Author(s) | David Gilland, Benjamin Welch, Seung Joon Lee, Brian Kross, Drew Weisenberger | ||||
Publication Date | October 2016 | ||||
Document Type | Meeting, Proceedings | ||||
Primary Institution | Thomas Jefferson National Accelerator Facility, Newport News | ||||
Affiliation | Exp Nuclear Physics / Technical Support Groups / Detector | ||||
Funding Source | Nuclear Physics (NP), Commonwealth of Virginia CIT grant | ||||
Proprietary? | No | ||||
This publication conveys | Technical Science Results | ||||
Document Numbers |
|
||||
Associated with an experiment | No | ||||
Associated with EIC | No | ||||
Supported by Jefferson Lab LDRD Funding | No |
Meeting / Conference
Meeting Name | 2015 IEEE NSS/MIC |
Meeting Date | 10/31/2015 |
Document Subtype | Paper (proceedings only) |
Proceedings
Title | NSS/MIC |
Editor(s) | |
Publisher | IEEE |
Refereed | No |
Page Number | 49 |
Year | 2016 |
Attachments/Datasets/DOI Link
Document(s) |
2015-ieee-nss-mic.pdf
(STI Document)
A_prototype_gamma_tomosynthesis_system_for_molecular_breast_imaging.pdf
(Accepted Manuscript)
|
DOI Link | |
Dataset(s) | (none) |